Section 1240. Radiation survey instruments  


Latest version.
  • A. The licensee or registrant shall keep sufficient calibrated and operable radiation survey instruments at each location where sources of radiation are present to make the radiation surveys required by this part and by Part IV (12VAC5-481-600 et seq.) of this chapter. Instrumentation required by this section must be capable of measuring a range from 0.02 mSv (2 mrem) per hour through 0.01 Sv (1 rem) per hour.

    B. The licensee or registrant shall have each radiation survey instrument required under subsection A of this section calibrated:

    1. At energies appropriate for use and at intervals not to exceed six months or after instrument servicing, except for battery changes;

    2. For linear scale instruments, at two points located approximately one-third and two-thirds of full-scale on each scale; for logarithmic scale instruments, at mid-range of each decade, and at two points of at least one decade; and for digital instruments, at three points between 0.02 and 10 mSv (2 and 1000 mrem) per hour; and

    3. So that an accuracy within plus or minus 20% of the true radiation dose rate can be demonstrated at each point checked.

    C. The licensee or registrant shall maintain records of the results of the instrument calibrations in accordance with 12VAC5-481-1410.

Historical Notes

Derived from Volume 22, Issue 25, eff. September 20, 2006; amended, Virginia Register Volume 24, Issue 18, eff. June 12, 2008.

Statutory Authority

§ 32.1-229 of the Code of Virginia.